ECCOS Mechanical Reprocessing
The solution for storage and mechanical reprocessing of your Aesculap power systems
Basic requirements for mechanical reprocessing in combination with the storage system:
- A suitable cleaner must be used (pH <11 for aluminum components)
- The temperature in the cleaning cycle must not exceed 60°C
- Clean for at least 3 minutes
- Apply intermediate rinse for at least 1 minute
- Carry out intensive final rinse with distilled, demineralized or fully desalinated water
- Only thermal disinfection: 5 minutes rinse at 96°C with distilled, demineralized or fully desalinated water
- Complete the program with a drying phase of at least 10 minutes at a temperature of not more than 120°
Additional Info Print
The solution for storage and mechanical reprocessing of your Aesculap Power Systems
Basic requirements for mechanical reprocessing in combination with our storage system:
> A suitable cleaner must be used (pH < 11 for aluminum components)
> The temperature in the cleaning cycle must not exceed 60°C
> Clean for at least 3 minutes
> Apply intermediate rinse for at least 1 minute
> Carry out intensive final rinse with distilled, demineralised or fully desalinated water
> Only thermal disinfection: 5 minutes rinse at max. 96°C with distilled, demineralised or fully desalinated water
> Complete the programme with a drying phase of at least 10 minutes at a temperature of not more than 120°C