ECCOS Mechanical Reprocessing

The solution for storage and mechanical reprocessing of your Aesculap power systems

Basic requirements for mechanical reprocessing in combination with the storage system:

  • A suitable cleaner must be used (pH <11 for aluminum components)
  • The temperature in the cleaning cycle must not exceed 60°C
  • Clean for at least 3 minutes
  • Apply intermediate rinse for at least 1 minute
  • Carry out intensive final rinse with distilled, demineralized or fully desalinated water
  • Only thermal disinfection: 5 minutes rinse at 96°C with distilled, demineralized or fully desalinated water
  • Complete the program with a drying phase of at least 10 minutes at a temperature of not more than 120°

Additional Info Print

The solution for storage and mechanical reprocessing of your Aesculap Power Systems
 

Basic requirements for mechanical reprocessing in combination with our storage system:
> A suitable cleaner must be used (pH < 11 for aluminum components)
> The temperature in the cleaning cycle must not exceed 60°C
> Clean for at least 3 minutes
> Apply intermediate rinse for at least 1 minute
> Carry out intensive final rinse with distilled, demineralised or fully desalinated water
> Only thermal disinfection: 5 minutes rinse at max. 96°C with distilled, demineralised or fully desalinated water
> Complete the programme with a drying phase of at least 10 minutes at a temperature of not more than 120°C