Infuvalve®

Accessory for infusion therapy to prevent back flow into infusion lines

An effective technical precaution to raise safety to complex infusion systems

  • Normally closed valve, to be opened with low pressure
  • Prevents backflow of fluid/blood into IV infusion line
  • Performs against over-and under-dosage as well as against bolus infusions
  • Maintains alarm function of pumps

Features

  • Functioning in all positions
  • PVC- / DEHP-free, Latex-free
  • Performance data
    • Minimum closing flow ≤ 0.1 ml/h
    • Opening pressure ≤ 20 mbar
    • Flow rate > 90 ml/minute G 40%
    • Flow rate reduction = 18 %
    • Burst pressure 6 bar

Note

  • Available as built-in valve in patient connector of IV set and as single packed valve
  • Infuvalve® does not prevent air entrainment - therefore must be capped
  • Infuvalve® does not retain microorganisms!